A pre‑slit septa is designed to improve needle penetration and maintain consistent sample integrity in vial‑based LC workflows. The pre‑slit opening provides a controlled point of entry that:
- Reduces insertion force, especially for thicker or blunter autosampler needles
- Minimizes septum coring, which can contaminate samples or block needles
- Decreases mechanical stress on the autosampler's injection mechanism
- Limits pre‑injection evaporation, which helps maintain accurate analyte concentration over time
Preventing Vacuum Formation During Sample Withdrawal
In many autosampler designs—such as Waters™ systems and other platforms with tight needle‑to‑septum interfaces—using a non‑slit septa can unintentionally create a partial vacuum in the vial during aspiration. This occurs when:
- The needle forms an airtight seal with the septum
- The sampler withdraws liquid faster than air can enter the vial
-
A negative internal pressure develops, leading to:
- Over‑aspiration (pulling more liquid than intended)
- Inaccurate injection volumes
- Potential needle damage or system errors in severe cases
A pre‑slit septa eliminates this issue by allowing a controlled amount of air to enter the vial as the needle withdraws fluid. This equalizes pressure and ensures the injection system can deliver:
- Accurate, reproducible sample volumes
- Proper pump and needle mechanics
- Stable performance across multiple injections, especially in high‑throughput sequences
Why It Matters
Proper septa selection is a critical but often overlooked variable in LC method robustness. Using a compatible pre‑slit septa helps ensure:
- More reliable quantitation
- Longer autosampler hardware life
- Reduced variability in multi‑run sequences
- Fewer injection‑related errors or instrument flags